DIN EN ISO 11608-5 Testing Autoinjectors

The DIN EN ISO 11608-5 standard describes the testing of autoinjectors. ZwickRoell offers testing systems for automated measurement of various characteristic values of autoinjectors.

Name Type Size Download
  • Industry brochure: Medical PDF 6 MB

Objective of the autoinjector test to DIN EN ISO 11608-5

An autoinjector is a medical instrument (medical device) with which liquid medication can be administered, and is therefore subject to strict quality controls. In the past, autoinjectors were mainly used when a drug needed to be administered as quickly as possible by medically untrained persons. Examples of this are adrenalin injectors used in cases of anaphylactic shock and morphine or atropine injectors used by soldiers in combat missions. Today there are two other factors that contribute to the increased use of autoinjectors.

One is injection vs. ingestion, meaning certain substances (for example, biopharmaceuticals) cannot be administered in tablet form, because they would be digested in the patient's gastrointestinal tract and would not enter the bloodstream. The other is that the autoinjector functions as an extension to the doctor's arm. Injectors guarantee a high level of application safety.

The correct injection and proper dose of medicine are vital to achieve the best therapeutic success. Therefore, pharmaceutical manufacturers strive to achieve a high level of automation in autoinjector technology. The patient simply removes the safety cap, positions the injector, and injects the drug by pressing a button. The process that follows is completely automated. However, this also means that all of the injector's relevant functions must be checked before the production batch is released on the market. This test is performed to DIN EN ISO 11608-5.

Autoinjector testing to DIN EN ISO 11608-5 with ZwickRoell

ZwickRoell offers testing systems that can perform the following automated tests:

  • Removal force of the safety cap
  • Autoinjector activation force and travel
  • Injection time measurement
  • Determination of the administered drug volume, including the last drops with an integrated high-resolution scale
  • Effective needle length for injection
  • Safety function of the needle shield
  • Testing of injectors with/without actuating button
  • Color recognition on autoinjector/actuating button
  • Acoustic recognition of start/end of injection
  • Mistake-proofing through design (Poka-Yoke)
  • Daily checks on load cell, scale, microphone, needle-length and fluid-presence sensors, as well as color recognition
  • Measurement of ambient humidity and temperature

The integrated test offers the benefit that all tests can be performed on a single specimen without changing test tools. The testing system can be expanded to accommodate additional tests, satisfying a variety of market requirements and product developments. With an optional microphone to provide audible feedback, the system can detect autoinjector clicks, which indicate the start and end of the injection. In addition, an HD camera can be integrated to document liquid emission. You can save these images with the test results to support traceability. The machine can be equipped with a monitoring mechanism that complies with the requirements of DIN EN ISO 13849. testXpert III testing software manages the control and evaluation of these additional functions, as well as the entire test control sequence.

    Summary of advantages

    • Multifunctional testing machine for all test sequences on one specimen
    • Elimination of operator influences through sequential tests on one injector
    • Five different Daily Checks guarantee reliable test results
    • Testing machine for autoinjectors with and without actuating button
    • Modular, flexible robotic system for fully automated specimen feeding
    • Fast cycle times
    • Validation support through expanded ZwickRoell qualification documentation (DQ/IQ/OQ)

    DIN EN ISO 11608-5 – Automated testing of autoinjectors

    When dealing with large production batches, a testing process that saves time and resources is often important. ZwickRoell offers scalable automation solutions for a wide range of requirements.

    The testing assistant roboTest N supports the user with simple applications. Typically, the testing of 30 autoinjectors can be automated with one magazine filling.

    The system can be easily and flexibly adapted to changing test requirements without requiring any special programming knowledge.

    The fully automated robotic testing system roboTest R performs the complete test sequence from removal of the cap, through testing of the autoinjectors, to disposal of the tested specimens and good/bad sorting. To ensure reliable and comparable test results, the measuring equipment is automatically verified on a daily basis (Daily Check). Due to using several autoinjector testing systems, even a large volume of specimens is processed quickly and reliably by roboTest R.

    We look for and find the optimal testing solution for every one of your requirements.

    Get in contact with our industry experts.

    We look forward to discussing your needs.

     

    Contact us

    Related products for autoinjector testing to EN ISO 11608-5

    Traceable, tamper-proof test results in accordance with FDA 21 CFR Part 11

    Traceable tamper-proof test results
    • Ever-increasing demands are placed on software used in the medical and pharmaceutical industries to document the traceability of completed actions.
    • With the traceability option, testXpert III enables logging of all actions and changes before, during and after the test, making test results and the documentation traceable and protecting them from manipulation.
    • Integrated user management and functions such as electronic records and electronic signature ensure that test results are always protected from tampering.
    • Together with the organizational measures and procedure instructions that apply to the individual companies themselves, the requirements of FDA in 21 CFR Part 11 are fulfilled.
    • ZwickRoell also offers a qualification service package (DQ/IQ/OQ) for validation support.
    • testXpert III logs all test and system related actions and settings and can therefore always answer the question “When does who do what, why and who is responsible?” 

    Learn more about the testXpert III option