- Metals
- 100,000 joules
- Drop weight test
- API-RP 5L3
- DIN EN 10274
- ASTM E436
High-energy drop weight tests
ZwickRoell DWT high-energy drop weight testers are used for testing and evaluating fracture surfaces of ferritic steels. The test method is described as a drop weight test in the following standards: API-RP 5L3, DIN EN 10274 and ASTM E436.
Specimens from the pipe wall of large oil and gas pipelines are subjected to the drop weight test to API 5L. The specimens with a height equal to the original wall thickness and widths of several centimeters are subjected to abrupt loading via a vertically falling weight with a striking edge. The energy (weight and release height) is set so that the specimen breaks at the cold-pressed notch and the fracture surface can be assessed visually.
The DWT high-energy drop weight testers from ZwickRoell are available in the following versions: 20, 30, 40, 60, 80 and 100 KJ. The product portfolio also includes drop weight testers for low-energy tests.
- The maximum drop height for all machine sizes is 5 meters by default. The high-energy drop weight testers DWT 20 kJ to DWT 40 kJ are alternatively available with a maximum drop height of 2.5 meters.
- The test tools are fitted outside the test area.
- The specimens are fed pneumatically within a time period of 10 seconds and meet ASTM E436 requirements.
Specimens for high-energy drop weight testers
Technical overview
| Model DWT | 20-2.5 | 20-5 | 30-2.5 | 30-5 | 40-2.5 | 40-5 | |
| Energy | 20000 J | 20000 J | 30000 J | 30000 J | 40000 J | 40000 J | |
| Item No. | 640064 | 640070 | 1093944 | 1072003 | 1086456 | 1093948 | |
| Load frame | |||||||
| Max. drop height | 2.5 | 5 | 2.5 | 5 | 2.5 | 5 | m |
| Drop weight | 820 | 410 | 1225 | 612 | 1630 | 820 | kg |
| Min. drop weight | 460 | 410 | 457 | 457 | 480 | 456 | kg |
| Additional single weights | |||||||
| 6.0 kg | - | - | 8 | 6 | 2 | - | piece(s) |
| 15.0 kg | 2 | - | 15 | 8 | 76 | 15 | piece(s) |
| Total weight | 6100 | 6350 | 6850 | 7050 | 7150 | 7050 | kg |
| Impact velocity | 7 | 9.9 | 7 | 9.9 | 7 | 9.9 | m/s |
| Lifting unit speeds | |||||||
| Rapid | 18 | 18 | 18 | 18 | 18 | 18 | m/min |
| Slow | 3 | 3 | 3 | 3 | 3 | 3 | m/min |
| Dimensions | |||||||
| A - overall height | 5900 | 8000 | 5750 | 8000 | 5900 | 7960 | mm |
| B - width | 3100 | 3060 | 3210 | 3060 | 3210 | 3060 | mm |
| C - width | 1140 | 1140 | 1140 | 1140 | 1140 | 1140 | mm |
| D - width | 1965 | 1965 | 1965 | 1965 | 1965 | 1965 | mm |
| Foundation dimensions | 2150 x 2150 | 2150 x 2150 | 2150 x 2150 | 2150 x 2150 | 2150 x 2150 | 2150 x 2150 | mm |
| Max. specimen dimensions | |||||||
| Length | 305 | 305 | 305 | 305 | 305 | 305 | mm |
| Width | 50 | 50 | 50 | 50 | 50 | 50 | mm |
| Height | 76 | 76 | 76 | 76 | 76 | 76 | mm |
| Option: Low-temperature cooler | |||||||
| Temperature | -80 | -80 | -80 | -80 | -80 | -80 | °C |
| Max. temperature variation | ±1 | ±1 | ±1 | ±1 | ± 1 | ±1 | °C |
| Number of specimens | 3 | 3 | 3 | 3 | 3 | 3 | piece(s) |
| Specimen dimensions | 50 x 76 x 305 | 50 x 76 x 305 | 50 x 76 x 305 | 50 x 76 x 305 | 50 x 76 x 305 | 50 x 76 x 305 | mm |
| Connected loads | |||||||
| Electrical supply: | |||||||
| At 50 Hz | 3x 400-440 | 3x 400-440 | 3x 400-440 | 3x 400-440 | 3x 400-440 | 3x 400-440 | V, N/PE |
| At 60 Hz | 3x 440-480 | 3x 440-480 | 3x 440-480 | 3x 440-480 | 3x 440-480 | 3x 440-480 | V, N/PE |
| Power consumption (full load), approx. | 11 | 7 | 11 | 5 | 11 | 5 | kVA |
| Compressed air | 8 | 8 | 8 | 8 | 8 | 8 | bar |
| Air consumption | 600 | 600 | 600 | 600 | 600 | 600 | l/min |
| Ambient temperature | +10 ... +35 | +10 ... +35 | +10 ... +35 | +10 ... +35 | +10 ... +35 | +10 ... +35 | °C |
| Humidity | 20 ... 90 | 20 ... 90 | 20 ... 90 | 20 ... 90 | 20 ... 90 | 20 ... 90 | % |
| Model DWT | 60-5 | 80-5 | 100-5 | |
| Energy | 60000 J | 80000 J | 100,000 J | |
| Item No. | 1065889 | 1094011 | 1070439 | |
| Load frame | ||||
| Max. drop height | 0.5 ... 5 | 0.5 ... 5 | 0.5 ... 5 | m |
| Drop weight | 1225 | 1630 | 2040 | kg |
| Min. drop weight | Approx. 831 | 831 | 856 | kg |
| Additional single weights | ||||
| 11.0 kg | 6 | 2 | - | piece(s) |
| 13.5 kg | - | 2 | - | piece(s) |
| 16.5 kg | - | - | 2 | piece(s) |
| 20.5 kg | - | 2 | - | piece(s) |
| 27.5 kg | 12 | 26 | 42 | piece(s) |
| Total weight | 9300 | 9800 | 12300 | kg |
| Impact velocity | 3.1 ... 9.9 | 3.1 ... 9.9 | 3.1 ... 9.9 | m/s |
| Lifting unit speeds | ||||
| Rapid | 18 | 18 | m/min | |
| Slow | 3 | 3 | m/min | |
| Dimensions | ||||
| A - overall height | 8000 | 8000 | 8580 | mm |
| B - width | 3060 | 3060 | 3060 | mm |
| C - width | 1360 | 1300 | 1360 | mm |
| D - width | 2500 | 2500 | 2500 | mm |
| Foundation dimensions | 2550 x 2550 | 2550 x 2550 | 3050 x 3050 | mm |
| Max. specimen dimensions | ||||
| Length | 305 | 305 | 305 | mm |
| Width | 50 | 50 | 50 | mm |
| Height | 76 | 76 | 76 | mm |
| Option: Low-temperature cooler | ||||
| Temperature | -80 | -80 | -80 | °C |
| Max. temperature variation | ±1 | ±1 | ± 1 | °C |
| Number of specimens (quantity by dimension) | 10 at (10x77x305) | 10 at (10x77x305) | 10 at (10x77x305) | piece(s) |
| Connected loads | ||||
| Electrical supply: | ||||
| At 50 Hz | 3x 400-440 | 3x 400-440 | 3x 400-440 | V, N/PE |
| At 60 Hz | 3x 440-480 | 3x 440-480 | 3x 440-480 | V, N/PE |
| Power consumption (full load), approx. | 11 | 11 | 11 | kVA |
| Compressed air | 8 | 8 | 8 | bar |
| Air consumption | 205 | 205 | 205 | l/min |
| Ambient temperature | +10 ... +35 | +10 ... +35 | +10 ... +35 | °C |
| Relative humidity (non-condensing) | 20 ... 90 | 20 ... 90 | 20 ... 90 | % |
Accessories for Drop Weight Tests